ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
358
000000-2071-464
10/2014 V_01
•
Apply drift correction. In the
Channel alignment
Processing function select the reference time series as
input 1
and the original time series as
input 2
. Check
the
Fit
box and the
Input 2
box and leave
Marker
box unchecked (Fig. 510). Select
lateral
from the
correction mode selection
drop down menu. Press
Apply
. The resulting image will have a doubling in
channels. The first half represents the original data
set, the second half the drift corrected data set
(Fig. 511). The number of channels depend on the
number of channels of the original time series.
•
Extract drift corrected channels. In the Processing tool
Copy
choose
Subset
. Select image with original and
drift corrected channels. Select under
Channels
the
second half of channels. Press
Apply
to obtain the
final drift corrected image (without the original non-
corrected channels).
Fig. 510
Methods Parameter of the
Channel Alignment
Processing tool
Fig. 511
The boxed channels represent the drift corrected ones.