Press
Continuous
to start a continuous scan.
Press the
Line Scan
•
The diagram (Fig. 309) shows a line with a
bump, which is the result of reflected light on
the optical axis not focused yet on the lower
glass surface You might adjust the line to lower
or higher values by altering the gain or laser
power.
•
Move the objective cautiously upward by
turning the Z focus knob of the microscope
stand. A lens moving upward will be indicated
by increasing numbers in the position field of
the
Z (Focus)
panel window.
•
If the focus position approaches the lower cover
slip surface, the position of the line will shift to
higher values (Fig. 310).
Fig. 308
Selection of Line Scan diagram
Fig. 309
Selection of Line Scan diagram
Fig. 310
Line diagram near the lower cover
slip surface