ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
126
000000-2071-464
10/2014 V_01
5.2.10
Acquisition Parameter – Focus
If the
Focus
tool tab is expanded the
Focus
control
tool becomes available (Fig. 168).
The
Z-Position
display box shows the current
z-position.
The number can be edited by typing or using the
arrowheads. The nosepiece or stage will automatically
update the new position.
By pressing the
Stop
button, you can stop the current
movement. The
Step Size (µm)
display box shows
the currently used step size for focus movement. The
value can be edited by typing or using the
arrowheads.
The
Home
button sets the z-position to the
predefined zero value.
Pressing
Load
will lower the nose piece (z-position)
for objective change.
Work
brings the z-position back to the actual used
position if the focus was zeroed or in
Load
position.
Home and Load positions can also be accessed from
the z-position display by pressing the
H
or
L
buttons.
The z-position can be altered by grapping and sliding the
plane
or by using the arrow slider. The
plane
shows the actual z-position.
Pressing
Manually
in the
Zero position
field will set the current z-position to 0.
If a
Z-Piezo stage
insert is in use and active then pressing
Leveling
in the
Piezo field
will center the
Piezo stage position in regard to the microscope focus drive.
The
Piezo Position (
µ
m)
input box allows the position of the Piezo to be altered.
The
Step Size (µm)
display box shows the currently used step size for focus movement of the
Z-Piezo
stage
.
Use of an optional high resolution focusing device (Piezo objective focus or Z-Piezo stage
insert)
The
Piezo Position
is indicated as a numerical value. This value and with it the focus position can be
changed. A change of this value updates the value of the microscope focus accordingly.
The Step Size, by which the Piezo focus moves when clicking on the arrows, can be set between
0.005 µm and 10 µm.
As soon as the focus position of the microscope is changed (via hand wheel or software), the piezo
objective focus is automatically levelled.
•
A click on the
Leveling
button moves the Piezo objective focus or Z-Piezo stage insert into the center
position of its travel range and the focus position of the microscope is reset to compensate for the
leveling. Therefore, the same Z-level remains visible (the current position is not set to zero).
The motor focus of the stand is operated in the same way via the relevant buttons. Moving into the
Work
or
Load
position is always performed via the motor focus and not via the Piezo objective focus.
Please see CHAPTER 4: ANNEX for further information on the Piezo objective focus and Z-Piezo
stage insert.
Fig. 168
Focus tool