Startup and Shutdown of the System
3.1
Startup of the System
The LSM system is equipped with a main power switch and two further switches labeled
System/PC
and
Components
that are
located at the front of the System Electronic Rack. The main switch has to be set
to
ON
to be able to switch the system on and off using the
System/PC
and
Components
switches. The
main switch can be used to switch off the complete system with one switch only. The electronics to run
the computer and the microscope are switched on with the
System/PC
switch. The laser module and the
scan head are switched on with the
Components
switch. These switches are also accessible via the
power remote witch (Fig. 5).
Refer to the operating manual and CHAPTER 4: ANNEX for handling of any Titanium-Sapphire-Laser used
for Multiphoton (NLO) Microscopy.
3.1.1
The Power Remote Switch
•
To start the system switch the main switch
(Fig. 5/
1
) to
ON
.
•
When set to
ON
the power remote switch
labeled
System/PC
3
) provides power to
the computer. This allows using the computer
and ZEN software offline.
The drives for floppy discs and CD/DVD of
the computer must not contain any data
storage item.
•
After switching on the computer type in the
user name and password to log on to the
computer.
•
After entries, confirm by clicking the
OK
button
or
Enter
.
−
The WINDOWS operating system desktop
appears on the screen, showing a number of
icons.
After complete startup, the computer is ready for
the components start.
•
To switch on the system completely put the
Components
switch (Fig. 5/
2
) also to
ON
. Now
the complete system is ready to be initialized
with the LSM Software.
1
Main switch ON/OFF
2
COMPONENTS switch ON/OFF
3
SYSTEM PC switch ON/OFF
Fig. 5
Power remote switch
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......