LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
237
If the kinetic will last longer than one day, a
warning is displayed (Fig. 317). Click
Ok
to
continue.
Define the following variables:
Time delay between consecutive data points = pi
(given start value = p0)
Start time of the measurement: pt,i
Number of data points: n
Data point counts = i (note: i starts with 0; so for
the first data point i = 0; or general i = n-1)
Summand/Factor = f
In
const
(constant) the period between two successive data points is constant and will be exactly the time
selected in the
Point to Point time
selection box.
With the definition of variables for
const
: p
i
= p
0
; p
t,I
= i x p
0
; i > 0
In
linear
the period between two successive data points as specified in the
Point to Point time
selection
box is increased linearly by a summand of which the initial value is defined in the
Summand
selection
box.
With the definitions of variables: p
i
= p
0
+(i-1) x f; p
t,I
= i x p
0
+ (i-1) x f; I > 0
In
exp
the period between two successive measuring cycles as specified in the
Point to Point time
selection box is increased by an exponential factor of which the initial base is defined in the
Factor
selection box. Note, that the first measurement is therefore assigned a 0 for the exponent.
With the definitions of variables: p
i
= p
0
x f
i-1
; p
t,i
= p
0
x (
Σ
f
i-1
); i > 0
Procedure to set up a kinetic
•
Enter the time spacing you want to start with in the
Point to Point time
input box of your kinetics
measurement.
•
Enter the number of data points in the
Number
box. Remember that the first data point is counted as
"0".
•
In the
Shape
selection box, set one of the shape options
const
,
linear
or
exp
.
•
Enter a number in the
Summand or Factor
selection box.
•
The
Use Kinetics
check box has to be ticked.
The system will display in the
Calculated Duration
box the total time of the experiment taken into
account measurement times, kinetic data points and the time the system needs for mechanical switches.
The start of the single data points can be retrieved from the
Save Info
panel.
The duration of the kinetic measurement will be displayed in the
Calculated Duration
display box
and in the
Time
selection area.
Fig. 317
Kinetic warning