CHAPTER 5 - MULTIPHOTON LASER SCANNING …
LSM 880
NLO (Non Linear Optics) Laser for LSM
ZEISS
10/2014 V_01
000000-2071-464
9
Horizontal shift of the image generated with Beam1 (NLO laser) to the right
(To shift the image horizontally to the left, just use the reverse direction buttons):
1.
Click on the right "fine movement" button of Beam1/Output/Axis2 repeatedly until an image
shift and illumination change is visible (image gets darker).
2.
Now click on the right "fine movement" button of Beam1/Input/Axis2 repeatedly until the image
is bright again.
-> The image generated by the NLO laser now has shifted a small distance to the right.
3.
Repeat steps 1 & 2 until the images of both beams are aligned horizontally.
If the offset is quite large, use the coarse buttons (not more than one click at once).