Pressing the
Select
button loads the active image
from the
Image Display
and displays it as Input
Image. Only Lambda stacks (only available in
LSM
imaging
modes) or multi-channel images can be
loaded.
•
In the
Linear unmixing
panel (Fig. 410), the number
of spectrally distinguishable fluorescent components
within the imaged sample can be selected from the
Components
selection box. The number of
extractable components cannot be higher than the
number of acquired channels. The ZEN software is
limited to a maximum of 10 components.
Fig. 410
Linear unmixing panel