ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
264
000000-2071-464
10/2014 V_01
(1)
Positioning in solution
•
Align a carrier in the
Carrier Position
tool. Focus into the solution.
•
Go to the
Measure
tool and the
Acquisition
panel. Choose
Current Position
or
Sample Carrier
and select the well(s) for the measurement.
(2)
Positioning in a cell
Before positioning you have to select appropriate cells and image those. An overview of fluorescent
expressing cells is best achieved by using the binoculars in transmitted and epi-fluorescent mode. Once
cells with adequate expression levels have been identified they can be imaged with the LSM.
Alternatively, suitable cells can be searched by using low zoom values in the LSM.
Visual inspection of the cells via binoculars
•
In the main toolbar, activate the
Locate
tab.
•
Open the
Microscope Control
tool (Fig. 352).
The
Microscope Control
tool permits motorized
functions including objective and reflector change
settings as well as the illumination mode like
transmitted or reflected light of the connected
microscope to be controlled via the software.
These microscope functions can also be
operated directly on the microscope stand
via assigned buttons or through the touch
screen, if available. In that case, any changes
are recorded by the software and displayed
in the relevant windows / panels.
For side and base port systems, the reflector
has to be on a none position.
Fig. 352
Microscope Control tool