Left Tool Area and Hardware Control Tools
LSM 880
184
000000-2071-464
10/2014 V_01
5.3.2.4
Selection of Laser Lines
•
Press the
Laser
button to display the laser
control expansion submenu (Fig. 265).
Alternatively you have access to the
Laser Lines
in the
Acquisition
tab. They provide settings
for:
−
Lasers
for laser power used in measurement
and pinhole adjustment,
−
Bleach
for laser power used in bleaching
Since the expansion submenu will close when
clicking in other areas, we recommend to use the
Laser Lines
from the
Acquisition
tab, if you want
to adjust the count rate.
Lasers
•
Press the
Visible Light
or
Invisible Light
tab to
display the laser control panel used for
measurements and pinhole adjustments (
Lasers
area) or bleaching (
Bleach
area) (Fig. 266). Press
again to close the panel.
•
Activate the lasers you want to use by checking
the corresponding boxes. The lasers with sliders
will appear. Set the requested laser attenuation
(%) using the sliders, the arrows, or typing in
the display field.
Note, that lasers for measurements and
(pre-)bleach can be selected independently
from each other.
If in
AOTF Dampening
a factor different
from 100 % was chosen, the set AOTF value
in
Lasers
and
Bleach
will be reduced by that
percentage.
Note, that lasers for measurements and pre-
bleach can be selected independently from
each other.
The same options are available for the
405 and 440 lasers and the NLO laser, if
available, in the
Invisible Light
laser
selection box (Fig. 267).
Fig. 265
Light Path tool with laser submenu
Fig. 266
Visible Light Laser panel
Fig. 267
Invisible Light Laser panel