If the defined free shape curve becomes too complicated or the selected
Scan Speed
is too high, the
following message appears in the status bar:
Maximum scanner acceleration exceeded!
•
In this case, reduce the
Scan Speed
set in the
Acquisition Mode
tool.
•
If the generated contour and the line traced by the scanner are not in coincidence, reduce the
Scan
Speed
by a further amount.
If no sufficient coincidence of the two lines can be achieved by the reduction of the scan speed, you have
to calibrate the scanner position signal.
Spot Mode
In the
Spot
mode
fluorescent or reflected light
occurring from a single voxel xyz is detected. In this
mode a spot can be defined by two perpendicular
lines in the
Image
using the
Spot Select
button
which is located in the lower part of the
Acquisition Mode
tool window (Fig. 139).
The following parameters for scanning a spot can
be defined:
Objective, Speed, Bit Depth and Offset in X
and Y (= position within the image)
These functions operate as described in section
In
Spot
mode it is not possible to acquire a
Z-Stack
. After definition of the spot position the
only possible scan mode is
Time Series
.
Spot scan data can also be displayed as images
consisting of two lines with 500 pixel per line.
When indicating the number of cycles in the Time
Series tool window, one cycle equals 1000 pixels.
Fig. 139
Acquisition Mode tool - Spot Mode