ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
378
000000-2071-464
10/2014 V_01
Repeat the process with the
Right
slider and input box; here the marking lines are in blue and the
illumination slider
in the
image control area
should be placed to 2, showing the right side
illumination of the original data.
The
Preview
in the image container will show the position of both sliders
Left
and
Right
as red and blue
lines respectively (Fig. 532).
To avoid a visible change in intensity along the two cut off lines, for
Mean Fusion
with
Fusion Subset
in X,
an overlap of both illumination sides should remain in the center of the image and the
Blending
slider should be used to define the number of pixels used within this overlap to crossfade both
illuminations sides.
The position of the sliders
Left
and
Right
as well as the
Blending
value must be defined for each View
individually.
Fig. 531
Dual Side Fusion – Fusion Subset in X
Fig. 532
Dual Side Fusion – Fusion Subset in X, Preview
and Image control area
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......