![Zeiss LSM 880 Скачать руководство пользователя страница 64](http://html2.mh-extra.com/html/zeiss/lsm-880/lsm-880_operating-manual_3433209064.webp)
ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
58
000000-2071-464
10/2014 V_01
(1)
Conventional setting of the microscope
Axio Observer.Z1
The Recording of microscope settings works as
described for the microscope Axio Imager.
For the conventional setting of the
Axio Observer.Z1, proceed as follows:
•
Click the
Locate
tab in the
Main
toolbar.
•
Place specimen on microscope stage.
−
The cover slip must be facing down.
•
In the
Objective
list box, select the required
objective.
•
Use the focusing drive (Fig. 59/
4
) to focus the
required specimen plane.
•
Select specimen detail by moving the stage in
X and Y via the XY stage fine motion control
(Fig. 59/
3
and
2
).
(2)
Microscope settings on Axio Observer for transmitted-light observation
•
Click on the
Transmitted light
button. Click the
On
button in the
Transmitted Light
panel and set
the transmitted light intensity via the slider or click on
3200 K
.
•
Click on the
Condenser
button and set the aperture via the slider in the
Condenser
panel. Set the
filter in the
Filter
selection box.
•
Click on the
Objective
button and select the objective by clicking on it.
•
Click on the
Reflector
button and select
None
.
•
Make sure there is no illumination from the reflected-light.
Fig. 58
Axio Observer.Z1 controls in the
Microscope Control tool
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......