Center Screen Area / Image Containers - Display and Image Analysis
LSM 880
530
000000-2071-464
10/2014 V_01
6.20.3
Obtaining an Anisotropy Image
Set up a multitrack configuration with alternating analyzers:
P
in track 1,
S
in track 2, leaving all other
settings equal (see Fig. 749). Assign track 1 to
Ch1-T1
, track 2 to
Ch1-T2
. Alternatively assign to Ch2-T1
and Ch2-T2 or ChS1-T1 and ChS1-T2, whatever physical channel you want to use.
Fig. 749
Channels tool with Ratio Type 1
formula selected for the Ratio1
channel using a G factor of 1.18
Fig. 750
Image calculator with anisotropy
formula and a chosen G-factor of
1.18 with absolute intensities
selected