LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
61
(1)
Microscope settings on Axio Imager for
transmitted-light observation
•
Set the reflector turret position to
None
and
click the
On
button for transmitted light.
•
Pull out the push rod (Fig. 61/
1
) to switch the
beam path of the tube in LSM position.
•
Control the brightness of the halogen lamp
with the potentiometer (Fig. 61/
4
) or the
Intensity %
slider in the
Microscope Control
tool.
•
Set the required transmission value of the gray
filters in the
Filter
frame.
•
Set the condenser and the luminous-field
diaphragm for KÖHLER illumination.
The laser-scanning mode automatically occludes
the halogen lamp even with
Transmitted Light
activated (
On
).
•
Click on the
Locate
tab in the Left tool area.
Press
Online
if needed.
•
Place specimen on microscope stage.
−
The cover slip must be facing up.
Fig. 60
Axio Imager.Z2 controls in the
Microscope Control tool
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......