LSM 880
Startup and Shutdown of the System
ZEISS
10/2014 V_01
000000-2071-464
23
3.2
Shutdown Procedure
Never shut down the computer by its main switch while the ZEN software is still active, or else
you will lose the currently set operating parameters and the images just scanned.
In the
Maintain
tab /
System Options
dialog window, activate
Laser off
on
Exit
in the
Shutdown
tab (section
). The Argon laser and Ti:Sa lasers will
automatically be switched off when you exit the LSM program.
3.2.1
Exiting ZEN Software
Fig. 13
Exit in the File menu