Center Screen Area / Image Containers - Display and Image Analysis
LSM 880
506
000000-2071-464
10/2014 V_01
N-FRET
or
Xia
method:
Xia and Liu,
Biophys J
81
, 2395 (2001)
NFRET = FRET 1/
√
(Dfd*Afa)
Displays an N-FRET image with intensities converted from the FRET index calculated for each pixel using
the Xia method. This method is similar to the Gordon method with the difference that for concentration
compensation the square root of donor and acceptor concentration is used. The resulting image is
properly corrected for variances in the fluorochrome concentration.
Parameter tab (Fig. 701):
Various Parameters of the image can be set in this
tab.
Prior to analysis the channels of the image need to
be assigned appropriately using the drop down
menu next to
Donor ch.:
,
Acceptor ch.:
and
FRET
ch.:
.
To determine the
Donor coefficient
values by
using a sample with only signal of the donor
fluorophore, one or more regions in the image are
defined and enabled as objects in the FRET tab. Of
this region the Donor coefficient values are taken
by clicking the
Analyze Donor
button.
The
Donor coefficient
include the Donor signal detected in the FRET channel/Donor signal detected in
the Donor channel (
Fd/Dd
)
and the Donor Signal detected in the Acceptor channel/Donor signal detected
in the FRET channel (
Ad/Fd
).
These values are saved for future use as a .fret file by choosing
Donor
in
the drop-down list and pressing afterwards the
Save
button.
To obtain the
Acceptor coefficient
values an image of a sample with only acceptor fluorescence is taken
and one or more regions are defined as objects and enabled in the FRET tab and the
Analyze Acceptor
button is pressed.
The values for the
Acceptor coefficient
include the Acceptor signal detected in the FRET channel /
Acceptor signal detected in the Acceptor channel (
Fa/Aa
), the Acceptor signal detected in the Donor
channel / Acceptor signal detected in the Acceptor channel (
Da/Aa
), the Acceptor signal detected in the
Donor channel/Acceptor signal detected in the FRET channel (
Da/Fa
). These values are saved for future
use as a .fret file by choosing
Acceptor
in the drop-down list and pressing afterwards the
Save
button.
The Gordon factor
G
depends on the fluorphores used and is part of the Gordon and Xia analysis. Please
refer to the original publications for further details (Gordon et al., Biophys J 74, 2702 (1998),
Xia and Liu,
Biophys J
81
, 2395 (2001)).
The total
Number of images
is set using the slider or by editing the field next to the slider, to choose
how many images of a time series shall be taken into account. Their values will appear in the table below
the image.
To load the saved Acceptor and Donor coefficient values for FRET analysis
Donor
or
Acceptor
are chosen
in the drop-down list respectively. Afterwards the
Load
button is pressed and the previously saved .fret
file chosen.
Fig. 701
FRET View Options Control Block,
Parameter tab, Last Analysis
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......