ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
124
000000-2071-464
10/2014 V_01
5.2.9.4
Channel Settings for Airyscan
The detection settings of the Airyscan can be
controlled in the
Channels
tool (Fig. 163).
The pinhole will be set automatically to a correct
opening depending on the Airyscan mode.
The effective pinhole is defined by the Airyscan
detection surface and its zoom optics.
The hardware pinhole will automatically be opened
at twice the value, to avoid any signal loss.
The effective pinhole values will be displayed in the
line below the Airyscan modes.
Detection gain is set in the same way as the
internal detectors by the Gain and digital gain
sliders.
SR, VP, CO, Pinhole
To set the system for increased resolution mode,
press the
SR
button (Fig. 164). In this mode, the
hardware pinhole has no effect and should not be
changed from the default position.
To set the system for virtual pinhole mode, press
the
VP
button (Fig. 165). In this mode, the
hardware pinhole has no effect and should not be
changed from the default position.
To set the system for conventional confocal mode,
press the
CO
button (Fig. 166). In this mode, the
hardware pinhole is effective and can be set at the
required value for confocal sectioning.
Fig. 163
Airyscan: Channels tool
Fig. 164
Airyscan Mode: SR
Fig. 165
Airyscan Mode: VP
Fig. 166
Airyscan Mode: CO
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......