LSM 880
Center Screen Area / Image Containers - Display and Image Analysis
ZEISS
10/2014 V_01
000000-2071-464
511
Fig. 706
Image window with two selected regions of interest in the intensity image
Correlation
allows you to select the type of
correlation analysis you want to perform. For a one
channel recording the only option is
Spatial auto-
correlation (S+A)
. For a two channel recording
you can select from the pull down menu between
Spatial auto-correlation (S+A)
and
Spatial
cross-correlation (S+C)
Channels
allow you to select the channel for
which a correlation should be performed. For a
one channel record auto-correlation setup, only
the active channel is available. For a two channel
auto-correlation setup you can select the channel
by the pull down menu (Fig. 708). The correlation
of the active channel will be computed and
displayed. Both channels will be displayed in the
Result
table and the correlation of the selected
row will be displayed.
Fig. 707
Correlation pull down menu
Fig. 708
Channels pull down menu for auto-
correlation