188 Keysight N9923-90001 User’s Guide
Data Analysis Features
All about Limit Lines
— Then
Options
.
— Then choose from the following:
—
Beep
——
On Pass
Audible beep sounds when a sweep passes.
——
On Fail
Audible beep sounds when a sweep fails.
——
Off
No audible beep sounds.
—
Warning On Off
Shows Pass or
Fail
indication on the display with each
sweep. Data points that fail are highlighted.
—
Limit Type
Then Fixed or Relative.
—
Offset
Shifts the limit line up or down. Enter an offset value using the
rotary knob, the
arrows, or by entering a value with the number
keys.
——Enter a
positive
offset to shift the limit line UP, usually used
with UPPER limits.
——Enter a
negative
offset to shift the limit line DOWN, usually
used with LOWER limits.
—
Margin
Shifts the limit line up or down like an offset, but only virtually.
The limit line does NOT appear to move. Enter a Margin value using
the rotary knob, the
arrows, or by entering a value with the
number keys.
Enter a
positive
margin to virtually shift the limit line UP.
Enter a
negative
margin to virtually shift the limit line DOWN.
How to Save and Recall Limits
In addition to saving and recalling limits with an *.sta file, limit lines can also be
saved and then recalled independent of the current mode or measurement.
Use this feature to recall and apply existing limit lines to measurements with
different settings or even a different frequency range.
Limit line files can be loaded into the FieldFox Data Link Software. Learn how in
the Help file that is included in the Data Link software.
— Press
Save/Recall
then
Select Device
to set the Device (Internal, SD card,
or USB flash drive) to be used for the Save or Recall. Learn more in
Type and Select Device” on page 196
— Then press
Limit
.
— Then choose from the following:
—
Save Limits
After limit lines have been defined, this saves the line
definition to a file on the specified device.
—
Recall Limits
Recalls a limit Line definition from the specified device.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...