54
Keysight N9923-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
How to set Output Power
— Press
Meas Setup 4
.
— Then
Output Power
—
High
Sets output power to the maximum achievable power at all
displayed frequencies. Output power is NOT FLAT across the
displayed FieldFox frequency span. Please see
for expected power levels.
—
Low
Sets output power to approximately –45 dBm, FLAT across the
displayed FieldFox frequency span.
—
Man
(default setting at -15 dBm) Set output power to an arbitrary
value, FLAT across the displayed FieldFox frequency span. If flattened
power can NOT be achieved, a warning message and beep occurs. To
achieve a flattened output power, reduce the power level or stop
frequency.
— Then press
Power Level
— Then enter a value using the numeric keypad, the
arrows, or the
rotary knob.
— Press
Enter
.
System Impedance (Z0)
To accurately view data presented in Smith Chart format, first set the System
Impedance.
Learn how to select Smith Chart format in
.
Learn how to make 75Ω measurements at the FieldFox Supplemental Online
Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
How to set System Impedance
— Press
Meas Setup 4
.
— Then
Settings
.
— Then scroll to System
Z0
and press
Edit
— Then type either
50
or
75
and press
Enter
Port Extensions
Port extensions allow you to electrically move the calibration reference plane
on either port 1 or port 2 after you have performed a calibration.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...