116 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
Sweeps in BOTH directions.
FULL 2-Port (QSOLT) – Mechanical Cal ONLY (Requires Option)
Quicker to perform QSOLT than Full 2-port. Corrects all S-parameters.
DUT: Insertable only
Standards: OPEN, SHORT, LOAD on port 1. Flush THRU between ports.
Recommended cal type for Insertable DUTs. A Cal Kit is NOT required for
port 2.
Sweeps in BOTH directions.
Enhanced Response Cal - Forward OR Reverse (May require an option)
Faster measurements than Full 2-Port.
DUT: Non-Insertable or Insertable. S-parameters PARTIALLY Corrected:
S21 and S11 (Forward)
OR
S12 and S22 (Reverse)
Standards: OPEN, SHORT, LOAD on
ONE
port. Defined THRU between
ports.
Sweeps in ONE direction.
TRL – Mechanical Cal ONLY
A complete 2-Port calibration with potentially better accuracy than Full
2-port. Corrects all S-parameters.
DUT: Non-Insertable or Insertable
Standards: Thru, Reflect, Line or variations of these. A TRL Cal Kit MUST be
selected to see this Cal Type
Sweeps in BOTH directions.
Learn more about TRL Calibration in the
Supplemental Online Help
:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
1-port Response Cals (Open or Short)
Calibrate the magnitude and phase of a measurement using ‘modeled’
mechanical standards. Measurement accuracy is better than Simple
Response Cals (available on the main Cal page - see
) but NOT as good as full 1-port cal. Corrects either S11 or S22.
Can be used with Isolation (see following section). Learn more in
DUT: Non-Insertable or Insertable
S-parameters Corrected: S11 or S22
Also select an Enhanced Response Optimization. Learn more in
Response Optimization” on page 119
.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...