Keysight N9923-90001 User’s Guide
107
Calibration for NA, CAT, and VVM Modes
—The connector choices that are offered depend on the frequency range.
For example, if you have selected a frequency range that exceeds 18
GHz, the only connector options you will be presented with in
QuickCal are 'other 50 ohm' or 'other 75 ohm'. This is because 3.5
mm connectors, which go beyond 18 GHz, are not supported with
QuickCal.
— For each DUT port:
—Select the connector type of the DUT using the
arrows or rotary
knob.
—Press
Change Gender
to change the gender of the DUT connector.
—Press
Next
… to make the selection.
—If present, press
Finish
.
—
Select/Change Cal Type
. The FieldFox always RECOMMENDS a Cal Type
that will quickly and accurately calibrate all of the displayed S-parameters.
The other Cal Types that are listed will NOT necessarily calibrate all of the
displayed S-parameters. Learn more about Cal Types in
.
—Using the
arrows or rotary knob, select a Cal Type, then press
Select
and
Finish
.
— Press
Start Calibration
No Connection
— Disconnect the DUT from the FieldFox.
— If a jumper cable or adapter is required to connect the DUT to the FieldFox,
connect those components to the FieldFox test ports connectors. The
effects of those components will be measured and removed during the
calibration, and only the effects of the DUT will be displayed in the
measurement results. These should be high-quality components!
— Do NOT connect anything at the cal reference plane where the DUT
connects.
— Press
Measure
Optionally Connect Load(s)
— Connect a Load standard at the point where the DUT will be connected and
press
Measure
under the following conditions:
—When a jumper cable has too much loss. If a ‘Cal Failed’ error appears
and you skipped the Load measurement, then perform the Cal again
but include measuring the Load.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...