Calibration for NA, CAT, and VVM Modes
—
Full 2-Port Cal
(default) Corrects all four S-parameters. Requires a
forward and reverse sweep, which causes slower trace
measurements. Learn why in
“Calibration Type” on page 114
.
—
Enhanced Response
Corrects forward (S21 and S11) and reverse (S12
and S22) measurements separately. Therefore, when measurements
in only one direction are required, this choice provides faster trace
measurements than a full 2-port cal. Also choose an Enhanced
Response Optimization. Learn more in
To find the best choice for your DUT:
— Press
Preset
then
Preset
.
— Select the appropriate S-Parameter and other settings (frequency range,
resolution, and so forth).
— Press
Trace 6
then
Math and Memory
then
Data->Mem
.
— Press
Data & Memory
.
— Press
Cal 5
then
More
— Select a CalReady Cal to compare with the current setting.
— Press
Esc
to exit the cal menu.
— View the differences in the two traces.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...