Calibration for NA, CAT, and VVM Modes
— If a jumper cable or adapter is required to connect the DUT to the FieldFox,
then connect
those
components to the FieldFox connectors. The effects of
those components will be measured and removed during the calibration,
and only the effects of the DUT will be displayed in the measurement
results.
These should be high-quality components!
— In NA, CAT, or VVM Mode, press
Cal 5
.
— Then
Mechanical Cal
This page summarizes the Mechanical Cal to be performed and allows you to
make changes. For best results, review the screen and make changes in the
following order:
1. S-parameters
– Verify that these are all of the S-parameters that you want
calibrated. If not, press
Esc
to terminate the calibration process, then
return to the Trace menu to display all of the traces to be calibrated. Learn
how in
“Multi-Trace Configurations” on page 43
2. DUT Connectors
– For each DUT port that is shown, verify the correct type
and gender of the DUT connector. If a DUT port is not shown, then that
port is not included in the list of S-parameters to be calibrated.
3. Cal Kit
- Verify the correct Cal Kit for each DUT port to be calibrated. See a
list of supported Cal Kits in the FieldFox Data Sheet. Find a link to the
online Data Sheet in
Appendix B:, “Specifications/Data Sheet,” on page
If the DUT connector type, gender, or Cal Kit is
NOT
correct for the DUT
ports to be calibrated, then:
—
Press
Change DUT Connectors
to select the correct connector types
that are on your DUT.
—
For each port:
—
Use the
arrows or rotary knob to change the DUT
connector type.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...