Keysight N9923-90001 User’s Guide
123
Calibration for NA, CAT, and VVM Modes
Calibration Method Summary
Calibration Method Summary
—
Mechanical Full 2-port Cal
is ALWAYS the most accurate Cal method. The
quality of a Mechanical Cal is completely dependent on the quality of the
OPEN, SHORT, LOAD standards and the quality of the standard
connections. Use the correct high quality standards to ensure the most
accurate calibration.
—
QuickCal
is ideal for DTF measurements. QuickCal should be performed
when good accuracy is required and when using a jumper cable or adapter.
— QuickCal is NOT recommended when the DUT has 3.5 mm-female or
SMA-female connectors.
—Even with the optional Load, phase accuracy begins to degrade when
the return loss is greater than about 20dB.
—
CalReady
is accurate ONLY when the DUT is directly connected at the test
ports and most accurate at room temperature.
—
For highest accuracy
, a new QuickCal or Mechanical Cal should be
performed:
—When the temperature changes more than about 10°F (5°C)
—When the connection to the DUT requires a different jumper cable or
adapters.
—When any of the following measurement settings change: Frequency
Range, Power Level, IF BW, and Resolution.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...