Overview
Overview
In normal NA Mode operation, the FieldFox measures the characteristics of a
test device as a function of frequency. With Time Domain (opt 010), frequency
information is used to calculate the inverse Fourier transform and display
measurements with time on the horizontal display axis. The response values
appear separated in time, allowing a different perspective of the test device's
performance and limitations.
The graphic below compares the same cable reflection measurement data in
both the frequency and time domain. The cable has two bends. Each bend
creates a mismatch or change in the line impedance.
The frequency domain of an S11 measurement shows reflections caused by
mismatches in the cable. It is impossible to determine where the mismatches
physically occur in the cable.
The Time Domain response shows both the location and the magnitude of each
mismatch. The responses indicate that the second cable bend is the location of
a significant mismatch. This mismatch can be gated out, allowing you to view
the frequency domain response as if the mismatch were not present. Learn
more about gating in
Markers that are created on a Time Domain trace can be used to pinpoint the
distance of the mismatch from the reference plane.
For more information on Time Domain theory, see
http://literature.cdn.keysight.com/litweb/pdf/5989-5723EN.pdf
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...