Keysight N9923-90001 User’s Guide 171
VVM (Vector Voltmeter) Mode - Option 308
2-Port Transmission Measurements
2-Port Transmission Measurements
A 2-Port Transmission measurement is used for measuring electrical length,
insertion loss, gain, or isolation of a DUT at a single CW frequency. The
FieldFox signal source is transmitted out the port 1 connector, through the
DUT, and into the port 2 connector. In the following image, the gain of an
amplifier is being measured.
Figure 11-3
2-Port Transmission measurement of an amplifier
How to make a 2-Port Transmission Measurement
1.
Press
Mode
then
VVM
then
2-Port Transmission
.
2.
Press
Freq/Dist
and enter the measurement frequency.
3.
Press
Meas Setup 4
then
More
then
Output Power
. Select
High
for passive
devices. Select Low for amplifiers.
4.
Calibrate the FieldFox using one of the methods described in
. Before performing a QuickCal or Mechanical
Cal, press
Meas Setup 4
then
Zero OFF
.
5.
Connect the DUT.
Notes:
— The magnitude value is the gain (positive value) or the loss (negative value)
of the DUT.
— The phase value is the difference in phase (in degrees) between the DUT
input and output.
— To measure isolation of the amplifier, reverse the connection to the amplifier
(
PORT 1
to the amplifier output).
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...