Keysight N9923-90001 User’s Guide 153
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
— Then
Trig Type
— Then choose from the following:
—
Freerun
The USB Power Sensor acquires data without waiting for a
trigger. This mode is NOT supported for Trace Graph measurements.
—
Internal
The USB Power Sensor acquires data when a valid pulse
signal is detected at the RF input. The following Trigger settings
determine the validity of the trigger signal.
—
External
The USB Power Sensor acquires data when a valid TTL
signal is detected at the USB Power Sensor external trigger input.
Trigger Edge
This setting determines which edge of a trigger signal initiates data
acquisition.
— Press
Sweep 3
— Then
Trigger
— Then
Trig Edge
— Then choose from the following:
—
Pos
Acquisition is triggered by the rising (positive) edge of a valid
signal.
—
Neg
Acquisition is triggered by the falling (negative) edge of a valid
signal.
Trigger Delay
After a valid trigger signal is received at the USB Power Sensor, data
acquisition begins after the specified Trigger Delay time PLUS the specified
Start time. See the Pulse Timing diagram in
“Trigger Settings Annotation” on
for more information.
— Press
Sweep 3
— Then
Trigger
— Then
Trig Delay
— Then enter a value using the numeric keypad, the
arrows, or the rotary
knob.
Auto Trigger
Used ONLY when Trigger Type = Internal, this setting determines whether the
trigger level is set manually or is set to the default level in the USB Power
Sensor firmware.
— Press
Sweep 3
— Then
Trigger
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...