154 Keysight N9923-90001 User’s Guide
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
— Then
Auto Trigger
— Then choose from the following:
—
ON
Trigger level is determined by the USB Power Sensor firmware.
—
OFF
The trigger level is set manually using the Trigger Level setting.
Trigger Level
Used when Trigger Type = Internal AND Auto Trigger = OFF (Manual).
— Press
Sweep 3
— Then
Trigger
— Then
Trig Level
— Then set the trigger level using the numeric keypad, the
arrows, or the
rotary knob. A line appears on
Tr 1
at the current trigger level setting. Data
acquisition occurs when the RF input signal crosses the line. This setting
can be used with either Positive or Negative edge triggering.
Trigger Settings Annotation
Trigger settings are annotated on the FieldFox screen as highlighted in
red
in
the following image:
In the previous graphic, the settings are: Trig Level = -10 dBm, Trig Type (Int),
Pos Slope (/), Delay
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...