118 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
Waveguide Cal Kits
Keysight sells two waveguide Cal Kit series: the premium 11644A series and
the economy N9911X series. Both are available online at
Effective Velocity Factor
Velocity factor is the speed at which an electromagnetic signal passes through
the transmission medium relative to the speed of light. This value is important
when distance is being calculated in DTF measurements (CAT mode) and Time
Domain (NA mode).
When the media is waveguide, the velocity factor changes with frequency.
FieldFox calculates this ‘effective’ velocity factor automatically. However, the
settings are different for CAT mode and NA mode.
CAT Mode - How to make Waveguide settings
These settings are necessary ONLY when making DTF measurements.
— Press
Measure 1
then select a
DTF
measurement.
— Press
Meas Setup 4
— Then
Settings
(Learn how to use the Quick Settings table in
— Set
Media
= Waveguide
. Frequency Mode =
BandPass
is automatically
selected for you.
— Scroll down to
Waveguide Definitions
. Select the
Waveguide Standard
being used. If your waveguide standard is NOT listed:
—Select
User Waveguide
. Then press
Done
.
—Then
DTF Cable Specifications
, then
Edit/Save/Recall Cables,
then
Edit Cable
.
—Scroll to set
Waveguide Definition
. The default setting is
VF Corr
=
Auto.
—Set the
Min
,
Max
, and
Cutoff Frequencies
.
—
Cable Correction = Auto
is the default setting. The Effective Velocity Factor
is calculated automatically based on the frequencies of the waveguide
standard. To override this setting, set Cable Corr =
Man
.
NA Mode - How to make Waveguide settings
These settings are necessary ONLY when your measurement requires electrical
delay or port extensions, or if using Time Domain Transform.
— Press
Meas Setup 4
— Then
Transform
— Then
Transform Settings
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...