Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings
Rx = 5.8 GHz to 6.8 GHz
Power Sensor Settling
The following two settings work together to allow you to choose the right
compromise between measurement speed and accuracy.
Each power sensor reading is considered “settled” when either:
— The difference between two consecutive readings are within the Tolerance
value.
—
Or
the Max (number of) Readings has been made.
The readings that were taken are averaged together to become the displayed
reading.
How to choose Power Sensor settling
— Press
BW 2
— Then
Tolerance
. Enter a value in dB between 0 and 10 using the numeric
keypad, the
arrows, or the rotary knob. Tolerance of 0 will never be
achieved, so the Max Readings will always be made. Press
Enter
— Then
Max Readings
. Enter a value between 1 and 20 using the numeric
keypad, the
arrows, or the rotary knob. Press
Enter
Sweep Settings
These settings provide added control of the FOPS measurement.
How to make additional sweep settings
— Press
Sweep 3
— Then choose from the following:
—
Power Setup
—
Source Enable
——
ON
Turn ON the internal source (required for FOPS).
——
OFF
Turn OFF the internal source
—
Src Nom Power
(-15 dBm default setting) Set output power to an
arbitrary value at the source CW frequency. Then enter a value using
the numeric keypad, the
arrows, or the rotary knob. Press
Enter
—
Num Points
or
Freq Step Size
Both settings control the number of power
sensor measurements between the Start and Stop frequencies. Enter one
setting and the other will be changed automatically. There must be at least
two data points.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...