104 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
— A
THRU
standard is used during some calibration steps to connect PORT 1
to PORT 2 in place of the DUT. A Flush THRU connection can be made
when cables that connect with the DUT can mate with each other. Learn
more in
. Otherwise, any reasonably short
cable can be used as a THRU standard.
Calibration Reference Plane
is the point (or points) at which the DUT and cal
standards are connected during a calibration. This can be at the FieldFox test
port connectors, or at the end of jumper cables or adapters.
CalReady
Every FieldFox contains a factory calibration that was performed at the port 1
and port 2 connectors, with -15 dBm input power, over the entire frequency
range of the FieldFox using a number of data points that allows reasonable
interpolation over the FieldFox frequency range.
This calibration, known as CalReady, allows you to immediately make
measurements of a DUT that is connected directly at the test ports (PORT 1
and/or PORT 2). CalReady corrects measurements when the FieldFox is turned
ON and when a measurement is created with no other correction in place.
When measuring a DUT using a jumper cable or adapter - NOT a direct
connection to a test port connector – then a QuickCal or Mechanical Cal is
recommended. CalReady can also be used to check the integrity of the jumper
cable that is attached to the test ports.
CalRdy
is shown when a measurement is corrected using CalReady.
You can change the properties of the CalReady calibration. Learn more in
Learn how to see when your factory CalReady calibration was performed in
“System Information” on page 235
.
How to Perform a Calibration
In CAT, NA, or VVM Mode, press
Cal 5
.
Press
Esc
at any time to end the calibration process.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...