Keysight N9923-90001 User’s Guide 141
Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings
——
Start
Annotated as Start and Stop frequencies.
——
Center
Annotated as Center and Span frequencies.
——Then choose from the following two methods to determine the
offset frequency
(the frequency that the receiver is offset from
the source). This is typically the frequency of the LO that is
used with the frequency converter. The frequency offset may
be positive, negative, or zero.
——
Freq Offset
Enter an Offset frequency. The receiver frequencies
are calculated.
——
Offset Calculator
Enter the Receiver Start and Stop
frequencies. Then press
Calculate Offset
.
——Press
Rx Sweep
to select the receiver sweep direction.
——
Fwd
Used for systems for which the output frequency is
Offs + Src. The output frequency of the DUT and the
receiver sweeps in the same direction as the source.
——
Rev
Used for systems for which the output frequency is
Offs - Src. The output frequency of the DUT and the
receiver sweeps in reverse direction as compared to the
source.
Invalid Settings
Violations of the following conditions will return an error message:
— The frequency span for the source and receiver must be the same.
— Both the source and receiver frequencies must be positive.
FOPS Configuration Examples
Negative Offset, RxSwp =
Fwd
(Offs + Src)
Src = 2 GHz to 3 GHz
Offset = -1 GHz
Rx = 1 GHz to 2 GHz
Positive Offset, RxSwp =
Rev
(Offs - Src)
Src = 2 GHz to 3 GHz
Offset = 3.8 GHz
Rx = 1.800 GHz to 800 MHz
Positive Offset, RxSwp =
Fwd
(Offs + Src)
Src = 2 GHz to 3 GHz
Offset = 3.8 GHz
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...