Keysight N9923-90001 User’s Guide
73
Time Domain – Option 010
Gating
Trace 4 shows the
frequency
response
with
gating enabled.
How to make Time Domain Gating settings
— Press
Measure Setup 4
— Then
Transform
— Then
Gating
— Then choose from:
—
On
Gating is being performed.
—
Off
Gating is NOT being performed.
Start, Stop, Center, and Span Gate Times
These settings specify the time in the trace to be gated (in or out). The gate
times can be specified using either Start and Stop or Center and Span.
— Press
Gating Start Stop
— A settings table appears that allows you to navigate and select the
following settings. These settings are made exactly like those made on the
Transform Settings table. Learn how in
Gating Type
This setting defines the type of filtering that will be performed for the gating
function. The gate start and stop flags on the display point toward the part of
the trace you want to keep.
Choose from the following:
—
Bandpass
KEEPS the responses within the Gating Start and Stop times.
—
Notch
REMOVES the responses within the Gating Start and Stop times.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...