128 Keysight N9923-90001 User’s Guide
Power Meter (USB) Mode
USB Power Meter Settings
USB Power Meter Settings
Select Power Meter Mode before making any setting in this section.
How to select Power Meter Mode
— Press
Mode_
.
— Then
Power Meter (USB Sensor)
.
Average / Peak
— Press
Measure 1
— Then choose from the following:
—
Average
Measures and reports the average of the power
measurements.
—
Peak
Available ONLY on the U202x power sensors. Measures and
reports peak power detected at sensor input.
Zeroing
The Keysight USB Power Sensors perform Internal Zeroing automatically.
Because Keysight USB Power Sensors have an internal switch, Internal Zeroing
does NOT require that the power source be turned OFF.
For highest measurement accuracy, when measuring power levels below –30
dBm, External Zeroing should be performed.
During External Zeroing, the power source MUST be turned OFF or the power
sensor disconnected from the power source (device under test). Power is OFF
when power sensor measurements are below –60 dBm. When power higher
than this is detected at the sensor, a warning message is shown on the
FieldFox and External Zeroing will not proceed.
How to perform External Zeroing
— Press
Cal 5
.
— Then
External Zero
.
While either Internal or External Zeroing is being performed, “Zeroing...” is
shown on the FieldFox screen. While this message is present, do NOT make
any changes to the FieldFox.
External Zeroing is NOT allowed on the U2020 X-series Power Sensors.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...