158 Keysight N9923-90001 User’s Guide
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Pulse Top
Pulse Top is NOT the same as the pulse Peak. Pulse Top is used to set and
report the reference power level. You enter a value in percent, and the Pulse
Top value is reported in the Auto Analysis table in dB down from 100% of the
reference power level.
How to set Pulse Top
— Press
Mrk ->/Tools
— Then
Pulse Top
— Then enter a percent value between 0 and 100 using the numeric keypad,
the
arrows, or the rotary knob.
— When set to 10%, 10 dB down from the 100% reference level is
reported.
— When set to 50%, 3 dB down from the 100% reference level is
reported.
— When set to 90%, 0.5 dB down from the 100% reference level is
reported.
Grid
Enables and disables the display of a 10 x 10 grid on the FieldFox screen.
How to enable Grid display
— Press
Mrk ->/Tools
— Then
Grid
—
OFF
Grid OFF
—
ON
A grid of X/Y lines appears on the screen.
Trace Memory
A memory trace is a pulse measurement
Tr 1
data trace (NOT the trace in the
Zoom window) that has been stored in the FieldFox volatile memory. It then
becomes static, meaning it is not updated with normal pulse measurements. It
can be displayed on the screen by itself, or with a pulse trace for comparison
purposes.
The
Tr 1
data trace is displayed in
yellow
.
The memory trace is displayed in
blue
.
How to display data and memory traces
— Press
Trace 6
.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...