Keysight N9923-90001 User’s Guide 155
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Marker Settings
Markers are not allowed with Meter-style measurements.
Markers are displayed on
Tr 1
and also on the zoom trace if the marker is
within the zoom window time span.
Up to six standard markers can be used to provide readouts of the Y-axis
amplitude at the current X-axis location.
The six standard markers can also be used as Delta Markers.
Learn more about standard markers in
“All about Markers” on page 174
Amplitude Markers
Amplitude markers are horizontal dotted lines that you can move to any
amplitude level on the screen. The two amplitude markers can be set
independently or used as a pair of delta markers.
Amplitude markers can be used as a visual comparison aid with the pulse
measurement traces.
How to make Amplitude Marker settings
— Press
Marker
— Then
More
— Then
Ampl Markers
— Then choose from the following:
—
OFF
Amplitude markers are NOT displayed.
—
ON
Amplitude markers ARE displayed at their previous positions.
— Then
Ampl Mkr 1 or 2
— Enter a value in dBm for the marker using the numeric keypad, the
arrows, or the rotary knob.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...