198 Keysight N9923-90001 User’s Guide
File Management
Saving and Recalling Files
— Only data traces are saved – not memory traces.
Corrections
— When parameters are corrected, data is returned even when not displayed.
Data is returned with the correction type that is applied for each
S-parameter. The correction type for each S-parameter is annotated in the
header.
— When parameters are NOT corrected or displayed:
—With Smith chart and Polar formats, Real/Imag zero-filled data
appears as (1.0, 0.0).
—With all other formats, LogMag/Phase zero-filled data appears as (0.0,
0.0).
> Display formats
In CAT Mode, S1P files are always Magnitude (dB) / Phase (angle) format.
In NA Mode:
— Smith chart and Polar display formats are ALWAYS saved in Real/Imag data
pairs.
— All other display formats are saved in LogMag/Phase (dbAngle).
> What Data Processing functions are included in saved SNP Data:
In NA Mode, the data processing functions (such as trace math) that are
included in the SNP data depends on a Preference setting. Click
System
, then
Preferences
, then
Preferences
, then scroll to
NA, SNP File Save
, then select
from the following:
—
Default
– SNP data does NOT include smoothing. SNP data saves are taken
AFTER data processing (location 2) on the processing chain.
—
Ad vanced
– SNP data includes the following:
—With Smith or Polar display format, SNP data saves are taken BEFORE
most processing (location 1 on the data processing chain).
—With any other display format, SNP data saves are taken just AFTER
Formatting and Smoothing is applied (location 3 on the data
processing chain). For each non-active trace, SNP trace data is
reprocessed using the data processing chain of the active trace.
See the FieldFox data processing chain at the
Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
Learn
more about System Preferences in
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...