244 Keysight N9923-90001 User’s Guide
Working with the Lithium-Ion Battery
Reconditioning Battery
Reconditioning Battery
Reconditioning a battery can be performed ONLY with the External Battery
Charger (N9910X–872).
1.
Insert a battery into the left bay of the external charger.
2.
Press the button labeled Push to Recalibrate Left Battery Bay
3.
The charger will charge the battery fully, discharge it completely, then
recharge it fully again. The entire process can take up to 12 hours.
4.
Install the battery into the FieldFox.
5.
On the System, Battery screen, verify that the battery is fully charged.
Battery Care
Visually inspect the battery periodically for signs of degradation, such as
swelling, cracking of the battery shell, or leakage of fluid. If degradation
occurs, replace the battery and dispose of it properly. Refer to
. Operation with a degraded battery could result
in damage to the FieldFox.
Maximizing Battery Life
— If AC power is not continuously available (Ex: in a field environment), use the
battery and recharge it when it holds a 20% to 50% charge, as shown on
the screen’s battery charge icon. At this charge level, 2 or 3 segments show
on the battery gauge.
— If the FieldFox is not going to be used for more than a week, remove the
battery. Best practice is to store the battery with a charge of about 50%. At
this charge level, 3 segments show on the battery gauge.
— Allow a battery to warm to room temperature before charging it.
Temperature shock can damage the battery chemistry and in some cases
cause a short circuit.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...