Keysight N9923-90001 User’s Guide
115
Calibration for NA, CAT, and VVM Modes
Definitions
:
Defined THRU
– A THRU standard for which there is a Cal Kit definition such as
in the 8551xA Cal Kits.
Flush THRU
(also known as Zero length THRU) – When the test port cables
mate together when measuring an Insertable DUT. No actual THRU standard is
required because the test port cables are directly connected. Flush THRU is
also a Defined THRU with a definition of zero loss and zero length.
Sweep Directions
– Both FULL 2-Port Cals listed below result in correction
that requires background measurements sweeps in both directions, regardless
of the displayed measurements. The displayed traces are updated at a slower
rate than Enhanced Response and 1-port calibrations, which require sweeps in
one direction only.
1-Port (OSL)
DUT: Not Applicable
S-parameters Corrected: S11 or S22
Standards: OPEN, SHORT, LOAD
Sweeps in ONE direction.
FULL 2-Port (Requires Option 122)
Most comprehensive calibration. Corrects all S-parameters
DUT: Non-Insertable or Insertable
Standards: OPEN, SHORT, LOAD on
BOTH
ports. Any THRU between ports.
For more information on the Unknown THRU process, see the FieldFox
Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
Non-insertable DUT – A device whose
connectors could NOT mate together. They
either do not have the same type of connector
or they have the same gender. This also means
that the test port cables could NOT mate
together as in the above diagram.
Insertable DUT – A device whose connectors
could mate together. They have the same type
of connector and opposite or no gender. This
also means that the test port cables could
mate together, as in the above diagram.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...