Keysight N9923-90001 User’s Guide 131
Power Meter (USB) Mode
USB Power Meter Settings
— Press
Scale / Amptd
.
— Then
More
.
— Then
Autoscale
.
Relative and Offset Power Measurements
Power measurements can be made that are relative to another reading or that
are offset by a fixed value.
— Use Relative to measure the difference between power levels.
— Use Offset to account for components or a long length of cable that is used
before the power sensor. For example, if using a 10 dB attenuator to protect
the FieldFox, enter an Offset Value of 10 to read the power as it would
appear without attenuation. If using a 10 dB amplifier, enter an Offset Value
of –10 to read the power as it would appear without amplification.
How to set Relative and Offset Power Measurements
— Press
Scale / Amptd
.
— Then choose from the following:
—
Relative Off On
When switched to ON, the FieldFox measures and
stores the current power level. Subsequent power measurements
display a power level which is relative to the stored value. When
relative is ON, the units change from dBm to dB or from Watts to %.
—
Offset Off On
Toggles ON and OFF the displayed power level using
the Offset value.
—
Offset Val
Value by which the power meter display is offset. A
positive value compensates for a component with loss. A negative
value compensates for a component with gain. Offset values are
limited to +/- 100 dB.
Display Units
Select whether to display power in dBm or Watts. When making relative power
measurements, these change to dB and %.
How to set Power Meter display units
— Press
Meas Setup 4
or
Scale / Amptd
,
More
— Then
Unit
.
— Then choose from the following:
—
dBm (dB)
Sets Power Meter display to dBm or dB.
—
Watt (%)
Sets Power Meter display to Watts or %.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...