182 Keysight N9923-90001 User’s Guide
Data Analysis Features
All about Markers
—
Peak Right
Moves the active marker to the next data point to the right that
meets the ‘Peak’ criteria. When no data points to the right meet the ‘Peak’
criteria, the marker does not move.
—
More
then
Min Search
(Not available in DTF) Moves the active marker to
the LOWEST amplitude data point of the trace.
The following applies to DTF Measurements ONLY:
—
More
—
Track Peaks ON OFF
When Track Peaks is ON, markers will find the new
peak with every sweep. While in this Mode, the markers can NOT be turned
OFF or moved.
—
Find 3 Peaks
Three markers are created and find the highest peaks (faults)
on the trace.
What Is a ‘Peak’
You define what the FieldFox considers a “peak” by setting the Peak
Threshold
and Peak
Excursion
.
These two peak criteria settings are set individually for each trace and for each
mode.
These settings do NOT survive a FieldFox or Mode Preset.
How to set Peak Criteria
— Press
Mrk ->/Tools
.
— In
CAT
and NA
modes:
—Then
Peak Search
.
— Then
Peak Criteria
— Then select the following:
—
Peak Threshold
Minimum amplitude (dB). To be considered valid, the
peak must be
above
the threshold level. The valley on either side can
be below the threshold level. Default = –190 for CAT and NA Modes;
–90 for SA Mode.
The Peak Threshold and Excursion settings are NOT available in DTF
measurements. To make these settings, create a non-DTF CAT mode
measurement, then change the measurement back to DTF.
CAT mode Loss measurements are displayed in positive values. In these
cases, the Peak Threshold setting is inverted. For example, Peak Threshold
= -20 will consider any peak that has less Loss than 20 dB to be valid.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...