Keysight N9923-90001 User’s Guide
81
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
— Then
Average N
where
N
is the current count setting.
— Enter a value using the numeric keypad. Enter 1 for NO averaging.
— Press
Enter
.
— While Trace Averaging is in process, press
Sweep 3
then
Restart
to restart
the averaging at 1.
Smoothing
Trace smoothing averages a number of
adjacent
data points to smooth the
peak-to-peak noise values on a displayed trace. The number of adjacent data
points that are averaged is known as the smoothing aperture. Aperture is set
by specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing can be used with any CAT format.
When enabled,
Smo
appears on the FieldFox screen.
How to set Smoothing
— Press
BW 2
.
— Then
Smoothing ON OFF
.
— Then
Sm. Aperture
and enter a value between 0 and 25 (percent) using the
numeric keypad.
— Press
Enter
.
Single or Continuous Measure
This setting determines whether the FieldFox sweeps continuously or only once
each time the
Single
button is pressed. Use
Single
to conserve battery power
or to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
— Press
Sweep 3
.
— Then choose one of the following:
—
Single
Automatically sets Continuous OFF and causes FieldFox to
make ONE sweep, then hold for the next Single key press.
Hold
is
annotated in the upper left corner of the display when NOT sweeping,
and changes to an arrow --> while the sweep occurs
—
Continuous ON OFF
Makes continuous sweeps. This is the typical
setting when battery power is not critical.
You can also use
Run / Hold +/-
to toggle between Single and Continuous.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...