119
Calibration for NA, CAT, and VVM Modes
— Under
Transform Stimulus Settings
, set the
Start
and
Stop
frequencies to
those of the Waveguide.
— Set
Stimulus = Bandpass Impulse
— Press
Meas Setup 4
or
Back
— Then
Calibration Settings
— Set
Media = Waveguide
— Set
Cutoff Frequency
. This is the absolute minimum frequency of the
waveguide. This value must be less than the Start Frequency of the
Waveguide.
Enter the calculated
Effective VF
value into the Velocity Factor setting.
Enhanced Response Optimization
When Enhanced Response Cal Type is selected, either for one calibration or for
CalReady, this setting optimizes the calibration based on the type of DUT
being measured. See also: CalReady Properties in
.
This setting does NOT survive Preset.
— Press
Cal 5
then
More
— Then
Enh.Response
— Then choose from:
—
NonReciprocal
(default) An amplifier is a Non-Reciprocal device
because it has gain in the forward direction, and very high loss
(isolation) in the reverse direction. This choice provides the best
correction for non-reciprocal devices, and reasonable correction for
reciprocal devices.
—
Reciprocal
A reciprocal DUT is a device in which the insertion loss
through the device is equal in both the forward (S21) and reverse
(S12) directions. A cable is a reciprocal device. This choice provides
the best correction for reciprocal devices. However, S11
measurements on non-reciprocal devices will appear to have more
return loss than the non-reciprocal choice.
Interpolation *
Highest measurement accuracy is achieved when the frequency range or
resolution settings remain the same during the measurement as when the
FieldFox was calibrated. If these settings change after performing a calibration,
the FieldFox will interpolate the calibration so that VERY accurate
measurements continue to be made.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...