Front Panel Menu Reference
SCPI Command
Not available
To initiate the measurement trigger system:
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:ACQuire command. If a trigger occurs before the trigger system is ready for it, the trigger will
be ignored. You can test the WTG_meas bit in the operation status register to know when the
instrument is ready to receive a trigger after being initiated.
Front Panel Menu Reference
SCPI Command
Select
Measure\Control
.
The Trig state field indicates "Initiated".
To query the WTG_meas bit (bit 3):
If bit 3 is set in the query response, the WTG_meas bit is true, and the instrument is ready to receive
the trigger signal. Refer to
for more information.
The instrument executes one measurement acquisition each time a bus, pin,
transient, or level trigger command is received. Thus, it will be necessary to initiate
the trigger system each time a triggered measurement is desired.
Trigger the Measurement
The trigger system is waiting for a trigger signal in the initiated state. You can immediately trigger the
measurement as follows:
Front Panel Menu Reference
SCPI Command
Not available
To generate a measurement trigger:
Alternatively, if the trigger source is BUS, you can also
program a
or an IEEE-488 <get> command.
As previously discussed, a trigger can also be generated by a digital pin or a trigger out signal
(transient). If a digital pin is configured as the trigger source, the instrument will wait indefinitely for the
trigger signal. If the trigger does not occur, you must manually return the trigger system to the idle
state. The following commands return the trigger system to the idle state:
Front Panel Menu Reference
SCPI Command
Select
Measure\Control
.
Then select the Abort control.
Fetch the measurement
After a trigger is received and the measurement completes, the trigger system will return to the idle
state. Once the measurement completes, FETCh queries can retrieve the most recent measurement
data without initiating a new measurement or altering the data in the measurement buffer.
4 Using the DC Power System
132
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......