![Keysight Technologies N69100 Series Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218157.webp)
Parameter
Typical Return
<
"date"
>
String program data. Enclose string parameters in single
or double quotes.
<last cal date>
Enters the calibration date: CAL:DATE "12/12/12"
CALibrate:LEVel P1|P2|P3
CALibrate:LEVel?
Advances to the next level in the calibration. P1 is the first level; P2 is the second; P3 is the third.
Parameter
Typical Return
P1|P2|P3
(none)
Selects the first calibration point: CAL:LEV P1
l
Some calibration sequences may require some settling time after sending CAL:LEV but before
reading the data from the DVM and sending CAL:DATA.
CALibrate:PASSword <
password
>
Sets a numeric password to prevent unauthorized calibration. This is the same as the
password.
Parameter
Typical Return
<
password
>
A numeric value of up to 15 digits
(none)
Set a new password to a value of 1234: CAL:PASS 1234
l
If the password is set to 0, password protection is removed and the ability to enter calibration
mode is unrestricted. The as-shipped setting is 0 (zero).
l
To change the password: unsecure calibration memory with old code, then set the new code.
l
If you forget your password, refer to
l
This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter
Typical Return
(none)
(none)
Store calibration constants into non-volatile memory: CAL:SAVE
Keysight N69100 Series Operating and Service Guide
157
5 SCPI Programming Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......