If a digital pin is configured as the trigger source, the instrument will wait indefinitely for the trigger
signal. If the trigger does not occur, you must manually return the trigger system to the idle state. The
following commands return the trigger system to the idle state:
Front Panel Menu Reference
SCPI Command
Select
Transient\Control
.
Then select the Abort control.
When a trigger is received, the triggered functions are set to their programmed transient values.
When the triggered actions are completed, the trigger system returns to the idle state.
You can test the TRAN-active bit in the Operation Status register to know when the transient trigger
system has returned to the idle state.
Front Panel Menu Reference
SCPI Command
Select
Transient\Control
.
The Trig state field indicates "Idle".
To query the TRAN-active bit (bit 6):
If a bit value of 64 is returned in the query, the TRAN-active bit is true, and the transient action is NOT
complete. When the TRAN-active bit is false, the transient action is complete. Refer to
for more information.
Programming a Step Transient
l
l
l
Enable the step function
First, you must enable the step transient function.
Front Panel Menu Reference
SCPI Command
Select
Transient\Mode
.
In voltage priority mode select Voltage mode. In
current priority mode select Current mode. In the
dropdown list, select Step. Then press
Select
.
To enable the transient function, use:
or
Program the step Level
Use the following commands to program the triggered output step level. The output will go to this
level when the trigger is received. In the front panel menu, you can only program the step level based
on the priority mode that you are operating in - voltage or current priority.
4 Using the DC Power System
116
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......