![Keysight Technologies N69100 Series Скачать руководство пользователя страница 254](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218254.webp)
Disassembly
SHOCK HAZARD Only qualified, service-trained personnel who are aware of the
hazards involved should remove instrument covers. Always disconnect the power
cable and any external circuits before removing the instrument cover. Some circuits
are active and have power for a short time even when the power switch is turned off.
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling.
Component damage can occur at electrostatic discharge voltages as low as 50 V.
The following guidelines will help prevent ESD damage during service operations:
l
Disassemble instruments only in a static-free work area.
l
Use a conductive work area to reduce static charges.
l
Use a conductive wrist strap to reduce static charge accumulation.
l
Minimize handling.
l
Keep replacement parts in original static-free packaging.
l
Remove all plastic, foam, vinyl, paper, and other static-generating materials from the immediate
work area.
Tools Required
l
T10 Torx driver for cover and board disassembly
l
T8 Torx driver for P600 disassembly
l
Small flat bladed screwdriver for battery removal
Cover Disassembly
1. Turn off the power. Remove all cables from the instrument.
2. Remove the 7 flat-head screws located on the top, the 10 flat-head screws along the sides, and
the 3 pan-head screws on the back (see figure below). Place the screws in a container so that you
do not lose them.
3. Remove the instrument cover.
7 Service and Maintenance
254
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......