Capacitive Load Boundary for Constant Voltage (CV) Operation
The following figures show the boundary limitations for capacitive loads with equivalent series
resistance (ESR) for constant voltage operation. Operation below the marginal stability lines with the
bandwidth settings may result in output instability. There are no absolute limits for DUT
capacitance, as the stability is also a function of ESR. Refer to
Load Capacitance and Lead Inductance
for details.
Capacitive Load Boundary (N69172A units, Comp 0-2 )
Capacitive Load Boundary (N69173A units, Comp 0-2 )
1. The output is stable in this region
2. The output may overshoot in this region
3. The output may be unstable in this region
Keysight N69100 Series Operating and Service Guide
43
1 Quick Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......