Instrument Calibration
Voltage Programming and Measurement Calibration
Current Programming and Measurement Calibration
Introduction
SHOCK HAZARD, LETHAL VOLTAGES - Some models generate output voltages up to
2,000 VDC! Ensure that all instrument connections, load wiring, load and sense
connections are insulated using high-voltage wire. Always attach the safety cover
provided, so that no accidental contact with lethal voltages can occur.
The instrument features closed-case electronic calibration; no internal mechanical adjustments are
required. The instrument calculates correction factors based on input reference values that you set
and stores correction factors in non-volatile memory until the next calibration adjustment is
performed. This EEPROM calibration memory is not changed by cycling power or *RST.
Refer to the
Recommended Test Equipment and Setups
section for a list of the equipment and test
setups required for calibration. Also refer to the
section for information about
connecting the voltmeter, current shunt, and load. Additional information about calibration follows.
l
The correct password is required to enter the Admin menu, which contains the calibration function.
The password is pre-set to 0 (zero). You can change the password once calibration mode is
entered to prevent unauthorized access to the calibration mode. Refer to
for
more information.
l
When calibrating the unit using SCPI commands, most steps involve sending a *OPC? query to syn-
chronize with the power supply’s command completion before proceeding. The response from the
instrument must be read each time *OPC? is given. In some steps, it may take up to 30 seconds for
*OPC? to respond.
l
Once started, you must complete each calibration section in its entirety. As each calibration sec-
tion is completed, the instrument calculates new calibration constants and begins using them.
These constants are not saved in nonvolatile memory until a SAVE command is explicitly given.
Keysight N69100 Series Operating and Service Guide
235
6 Verification and Calibration
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......